  • <tr id='KCfeTt'><strong id='KCfeTt'></strong><small id='KCfeTt'></small><button id='KCfeTt'></button><li id='KCfeTt'><noscript id='KCfeTt'><big id='KCfeTt'></big><dt id='KCfeTt'></dt></noscript></li></tr><ol id='KCfeTt'><option id='KCfeTt'><table id='KCfeTt'><blockquote id='KCfeTt'><tbody id='KCfeTt'></tbody></blockquote></table></option></ol><u id='KCfeTt'></u><kbd id='KCfeTt'><kbd id='KCfeTt'></kbd></kbd>

    <code id='KCfeTt'><strong id='KCfeTt'></strong></code>

    <fieldset id='KCfeTt'></fieldset>
          <span id='KCfeTt'></span>

              <ins id='KCfeTt'></ins>
              <acronym id='KCfeTt'><em id='KCfeTt'></em><td id='KCfeTt'><div id='KCfeTt'></div></td></acronym><address id='KCfeTt'><big id='KCfeTt'><big id='KCfeTt'></big><legend id='KCfeTt'></legend></big></address>

              <i id='KCfeTt'><div id='KCfeTt'><ins id='KCfeTt'></ins></div></i>
              <i id='KCfeTt'></i>
            1. <dl id='KCfeTt'></dl>
              1. <blockquote id='KCfeTt'><q id='KCfeTt'><noscript id='KCfeTt'></noscript><dt id='KCfeTt'></dt></q></blockquote><noframes id='KCfeTt'><i id='KCfeTt'></i>

                测试技术

                ① 脉动流测试※技术

                脉动测试机能够模拟人体心脏系统的血液动力学,以及与此相关的血液粘度、动脉及心室压力,动脉流量等因素。通过软件,脉动测试¤机可以与控制箱之间实现信号接收及发◥送,从而能够记录/研究数据,调整驱←动波形,温度控制,位移、流⊙量及压力控制等。

                ② 稳态流测试√技术

                测试依据:ISO 5840-2:2015
                稳态前向流:在5 L/分钟-30 L/分◣钟的流量下以5 L/分钟的间隔,测量瓣膜流入流出侧的跨瓣压差。
                稳态泄漏:在5.2 kPa–26 kPa(40 mmHg–200 mmHg)的反@压下取5个等压差节点,测量瓣膜的泄漏情况。

                ③ 计算流体动力学

                使用粒子图像测速和计算流体动力学的方法确定瓣膜附近的流动特征。

                ④ 疲劳测试

                疲劳测试机用于测试瓣膜的耐久性。每个瓣膜均可单独调节并采用一对一的∏电机驱动来避免各测试瓣之间的干扰。